Troubleshooting Guide – Western blot

Immunochemical methods that use antibodies for the analytical determination of proteins are widely used in clinical diagnostics and drug development. The already traditional methods include Western blot, ELISA and RIA. In this post we want to share with you the “troubleshooting” for Western blot. We believe that it will be a kind of first aid if the experiments do not work as they should.

ProblemPossible sourceProposal
Weak / No staining Sample processingIncrease the amount of protein in the sample.Confirm by an alternative method that the protein is present in the sample.
Electrophoresis took too longShorten the electrophoresis time and check the dye progression by looking at the front of the gel.
Turn off the power when the front of the dye is close to the bottom of the gel.
Old or degradedprotein samples Test a new sample of the target protein.
Protein exited the membrane / protein was not transferred correctlyUse a membrane with a smaller pore size.
Reduce the stress for lower molecular weight proteins (less than 10 kDa). Make sure that the membrane is pre-equilibrated according to the manufacturer’s instructions.
Check and optimize the transfer time, as the appropriate transfer time varies with the MW of the target protein.
Make sure that the blotting cartridge is correctly assembled and that there is sufficient contact between the gel and the membrane. (Refer to the Western blot protocol for information on the correct composition of the blotting cartridge.)
Make sure that the correct transfer buffer is used.
The presence of SDS during transfer may reduce protein binding using a nitrocellulose membrane.
Add 20% methanol to transfer buffer to increase binding.
Check protein transfer by staining the membrane with a reversible dye such as Ponceau S to visualize the major bands.
The protein has a pl greater than 9Replace the pfur with a higher pH buffer (eg CAPS buffer, pH 10.5).
Protein overlaid with blocking bufferTest another blocking buffer.
Optimize the protein concentration in the blocking buffer. We recommend 3% skim milk powder in deH2O.
Presence of enzyme inhibitors (eg azide)Make sure that the buffers do not contain sodium azide as it would interfere with the HRP signal.
Old or degraded primary antibodyCheck manufacturer’s recommendations for storage conditions and / or expiration. If the expiration date has expired or the storage period has expired, purchase a new antibody.
If the antibody has been repeatedly frozen and thawed, purchase a new antibody, as the structure and ability to bind protein: antibody may be affected.
Perform a Dot Blot to ensure that the antibody is still active
Insufficient amount of primary and / or secondary antibodyCheck the manufacturer’s recommendations for antibody concentrations.
Test multiple concentrations to find the optimal conditions for your experiment.
Incorrect primary antibodyEnsure that the primary antibody reacts with the target protein from the organism being studied (eg Rabbit anti-human
VEGF -> human VEGF).
Incorrect secondary antibodyMake sure that the secondary antibody used is a primary antibody against the body (eg donkey anti-rabbit
IgG -> Rabbit anti-human VEGF).
Insufficient incubation with primary antibodyIncubation with antibody should last for at least 1 hour at room temperature or overnight at 4 ° C.
Insufficient incubation with secondary antibodyExtend the incubation time with the secondary antibody (see manufacturer’s recommendations for incubation time and temperature).
The membrane is washedShorten the washing time and / or reduce the number of washes.
Use a wash buffer that does not contain detergent or has a lower detergent concentration.

Incompatible developing agent
Make sure that the developing reagent used is compatible with the enzyme conjugate.

Ensure that the substrate is properly prepared according to the manufacturer’s instructions.
Excess saltReduce the amount of salt present in the wash buffer.

Reduce the amount of salt in the antibody dilution buffer.
Degraded / incorrect

developing agent
Make sure that the developing reagent has been prepared correctly.

Check that the reagent is still active. 
If not, purchase a new reagent.
The film is bad / expired for the ECL methodMake sure the film is not expired or exposed. 
If so, buy a new vilm.
For the ECL method, ECL developing solutions are old / expiredUse fresh developing ECL solutions.

With the ECL method, the film exposure time is too short
Increase the shutter speed of the movie.

With the ECL method, the reaction is disturbed by the foil
Use a different type of film.
High BackgroundAntibody concentration too highCheck the manufacturer’s recommendations for the concentration of primary and / or secondary antibodies.

Test different concentrations to find the optimal conditions for your experiment.
Incubation with antibody for too longCheck the manufacturer’s protocol to ensure that the correct incubation time and temperature are used. 
If the manufacturer did not provide this information, optimize the incubation time.
Non-specific binding of antibodiesEnsure that the primary antibody is specific for the target protein only.

Perform the control by incubating the membrane without the primary antibody. 
If bands develop, select another secondary antibody.

Select a secondary antibody from another organism.
Cross-reaction of antibodies with proteins in blocking buffer and / or sampleUse a different blocking buffer formulation.

Goat antibodies will react with BSA due to the slight recognition of bovine protein by the goat antibody. 
This will lead to a purple sheen on your membrane. 
If this happens, use an antibody from another species or a different blocking buffer.

If you suspect a secondary antibody from the problem, dilute it in a buffer containing 1-5% normal serum from the same organism as the target protein.

Inefficient blocking buffer
Use a different blocking buffer formulation.

Add detergent to a blocking buffer, such as TWEEN-20.
Insufficient blockingUse a different blocking buffer formulation.

The blocking buffer lot may be inactive. 
Use a new blocking buffer.

If you block overnight at 4 ° C, the blocking efficiency may be reduced due to the lower temperature. 
Try incubating for one hour at room temperature.

Add TWEEN-20 to blocking buffer.

Optimize the incubation time and / or temperature with blocking buffer.

insufficient washing
After each step, wash the membrane 3 times for 10 minutes. 
If this is not sufficient, increase the washing time.

Try another detergent in wash buffer or add TWEEN-20 if it is not already present.

Increase the volume of wash buffer used.

Incubation in staining substrate for too long
Shorten the incubation time in the staining solution. 
Observe development until bands appear and stop the reaction by washing the membrane in deH 2 O.
Excess enzymeReduce the concentration of the selected enzyme conjugate.
Interference caused by the presence of endogenous enzymesFat-free milk powder contains endogenous biotin and may interfere with the avidin / biotin detection system. 
Use a different blocking buffer, such as 3% BSA.

Contaminated reagents
Filter the buffers before use to remove contaminants. 
Prepare fresh solutions and repeat the western blot.
Membrane problemsTest new membranes.

Throughout the procedure, the membrane can only be dried between transfer and immunostaining. 
Once immunostaining of the membrane is initiated, the membrane cannot be dried until immunostaining is complete. 
If the membrane has dried during immunostaining, repeat the western blot.
With the ECL method, the ECL film is overexposedShorten the exposure time of the movie.

If the signal from the target protein is too strong, wait 5-10 minutes and expose again.
Stains backgroundUneven mixing during incubationEnsure that the membrane is shaken evenly during incubation by placing it on a rocker or shaker.

Insufficient incubation buffer
Ensure that a sufficient amount of solution is present during each incubation period to completely immerse the membrane and that the membrane floats freely in the solution.

Presence of air bubbles during transmission
Before starting the blot, gently remove any air bubbles between the layers of the blotting cartridge. 
You can use a clean glass rod or test tube.
Diaphragm problemsEnsure that the membrane is thoroughly moistened according to the manufacturer’s protocol.

Handle membranes very carefully. 
Incorrect membranes can cause non-specific binding.

Handling membranes with bare hands can lead to membrane contamination. 
Use tweezers and gloves when handling the membrane.

Aggregation of HRP conjugate
If using HRP, filter the conjugate to remove HRP aggregates.

Use fresh HRP conjugate.
Reagent contaminationFilter the buffers before use to remove contaminants.

Prepare fresh buffers and repeat the western blot.
Non-specific stripsAntibody concentration too highCheck the manufacturer’s recommendations for primary and / or secondary antibody concentrations.

Test multiple concentrations to find the optimal conditions for your experiment.
Old or degraded antibodyCheck manufacturer’s recommendations for storage conditions and / or expiration. 
If the expiration date has expired or the storage period has expired, purchase a new antibody.

If the antibody has been repeatedly frozen and thawed, purchase a new antibody, as the structure and ability to bind protein: antibody may be affected.
Non-specific binding of the primary antibodyReduce the concentration of primary antibody.

Check the protein concentrations added to the gel.

Use a more specific antibody, such as a monoclonal antibody or an affinity purified polyclonal antibody.

Add a small amount of detergent, such as TWEEN-20 (0.1-0.5%), to the primary antibody and / or blotting buffer solution.

Increase the number / time of washing.
Non-specific binding of secondary antibodyReduce the concentration of the secondary antibody.

Perform the control by incubating the membrane without the primary antibody. 
If bands develop, select another secondary antibody. Add a small amount of detergent, such as TWEEN-20 (0.1-0.5%), to the secondary antibody solution.

Increase the number / time of washing.

Target protein aggregation
Add or increase the concentration of a reducing agent such as DTT or BME to break up any disulfide bridges.

Before applying to the gel, heat the protein sample in a hot water bath for 5-10 minutes.

Check the protein concentration applied to the gel. 
Too high a concentration can cause protein aggregation. 
Repeat the Western blot with a lower protein concentration.
Protein degradationPrepare fresh samples and minimize the number of freeze-thaw cycles

.

Add protease inhibitor to samples before storage.

Store samples at -20 ° C to -80 ° C.
Insufficient blockingTest for different blocking buffer compositions. 
The blocking buffer lot may be inactive. 
Use a new blocking buffer.

If you block overnight at 4 ° C, the blocking efficiency may be reduced due to the lower temperature. 
Try incubating for one hour at room temperature.

Add TWEEN-20 to blocking buffer.

Optimize the incubation time and / or temperature with blocking buffer.
SDS causes non-specific binding

to immobilized protein bands
Ensure that the membrane is sufficiently washed after blotting.

Do not use SDS during immunostaining if you are using a nitrocellulose membrane.
Excess enzymeReduce the concentration of the selected enzyme conjugate.

Reagent contamination
Filter the buffers before use to remove contaminants.

Prepare fresh buffers and repeat the Western blot.
White stripes

(ECL visualization method)
The antibody or protein concentration is too high / excessive

signal generated
White streaks are the result of rapid substrate consumption due to the presence of too much conjugate. 
This is the result of high concentrations of antibodies and / or proteins. 
Check the manufacturer’s recommendations for antibody and / or protein concentrations.