EBV detection with Corbett Research

Inactivated EBV virus (NATtrol EBV, Zeptometrix)
was used in all experiments.

NATtrol EBV is prepared
from highly purified virus, isolated from cell culture,
which is chemically and enzymatically treated to alter
its surface proteins rendering it unable to infect a
host cell. While modified on its surface, the virus
still retains an intact and complete viral genome,
making it an ideal control1.
Purified viral DNA from NATtrol EBV was used
to optimize and validate the conditions for a
quantitative PCR (qPCR). Serial dilutions of
EBV DNA were tested in triplicate with custom
designed primers for the LMP-2A EBV gene
(Forward: 5’-CATGTTCTGGGTCAAAAGG and
Reverse: 5’-TGCTGATTCTGCTAACACTG).
PCR reaction: in a 25 μL reaction volume; 0.2 U Taq
Polymerase, 200nM primers, 1.5 mM MgCl2,
2 PD-WP-00028 Issue 2/2015-01
© 2015 DNA Genotek Inc., a subsidiary of OraSure Technologies, Inc., all rights reserved. www.dnagenotek.com • [email protected]
For research use only, not for use in diagnostic procedures.
Not available for clinical diagnostic use in the United States.
1 μM Syto®9 dye (Invitrogen), PCR Buffer without
MgCl2 (Roche), 0.1 mg/mL BSA. PCR conditions
were used: initial denaturation for 5 minutes at 95°C
(1 cycle); denaturation for 30 seconds at 95°C,
annealing 40 seconds at 58°C and extension for
30 seconds at 72°C (40 cycles), final extension for
7 minutes at 72°C. All qPCR runs were performed
on the Corbett Life Sciences Rotor-Gene™ and the
data was analyzed using the default quantitation
analysis with the built-in software, Rotor-Gene™ 6000
ver. 1.7. The qPCR assay resulted in a 4 log linear
dynamic range (80 – 20,000 copies cp of EBV virus
per reaction), a reaction efficiency of 1.01 ±0.07,
R2 of 0.98, no detectable amplification in the
no-template control and a single amplified product
with the expected theoretical melting temperature.